PET/SPECT Test Phantom

It is vital to test the response, reliability and to determine image artifacts produced by the nuclear medicine tomography systems due to some systemic limitations before their use in clinical studies. That cannot be reliably achieved by scanning patients. Different types of test phantoms are commer...

Full description

Bibliographic Details
Main Author: Shah, Sayed Inayatullah
Format: Conference or Workshop Item
Language:English
Published: 2010
Subjects:
Online Access:http://irep.iium.edu.my/6054/
http://irep.iium.edu.my/6054/1/IRIIE_2010_Poster.pdf
id iium-6054
recordtype eprints
spelling iium-60542011-11-21T21:22:10Z http://irep.iium.edu.my/6054/ PET/SPECT Test Phantom Shah, Sayed Inayatullah QC Physics It is vital to test the response, reliability and to determine image artifacts produced by the nuclear medicine tomography systems due to some systemic limitations before their use in clinical studies. That cannot be reliably achieved by scanning patients. Different types of test phantoms are commercially available with accurately/precisely known composition, which helps in acquiring prior knowledge about the capability and drawbacks of a nuclear medicine tomography system. A cold regions insert, that can be placed into a cylindrical tank of a commercially provided PET/SPECT phantom has been developed (for R A Carlson’s phantom - which consists, hot, cold and linearity inserts where the hot regions insert is comprised on a “V” shaped eight pairs of various sized holes with a cold background). The new produced cold regions insert, consists solid plastic rods, 5 cm long (each) with same diameters and exactly the same arrangement as hot region holes pairs insert of R A Carlson’s phantom. Both inserts have been attached with each other by aligning hot and cold regions for similar cross sectional views. The data acquired by this arrangement provides a simple way for measurement and comparison of the performance of a system i.e., image resolution in terms of position and size of hot and cold regions in the phantom, region(s) detectability. Also the accuracy of scatter, attenuation correction and image reconstruction techniques in nuclear medicine tomography may be tested. 2010 Conference or Workshop Item PeerReviewed application/pdf en http://irep.iium.edu.my/6054/1/IRIIE_2010_Poster.pdf Shah, Sayed Inayatullah (2010) PET/SPECT Test Phantom. In: IIUM Research, Invention and Innovation Exhibition IRIIE 2010, 26 - 27 January 2010, Culture Activity Centre, IIUM, Gombak. (Unpublished)
repository_type Digital Repository
institution_category Local University
institution International Islamic University Malaysia
building IIUM Repository
collection Online Access
language English
topic QC Physics
spellingShingle QC Physics
Shah, Sayed Inayatullah
PET/SPECT Test Phantom
description It is vital to test the response, reliability and to determine image artifacts produced by the nuclear medicine tomography systems due to some systemic limitations before their use in clinical studies. That cannot be reliably achieved by scanning patients. Different types of test phantoms are commercially available with accurately/precisely known composition, which helps in acquiring prior knowledge about the capability and drawbacks of a nuclear medicine tomography system. A cold regions insert, that can be placed into a cylindrical tank of a commercially provided PET/SPECT phantom has been developed (for R A Carlson’s phantom - which consists, hot, cold and linearity inserts where the hot regions insert is comprised on a “V” shaped eight pairs of various sized holes with a cold background). The new produced cold regions insert, consists solid plastic rods, 5 cm long (each) with same diameters and exactly the same arrangement as hot region holes pairs insert of R A Carlson’s phantom. Both inserts have been attached with each other by aligning hot and cold regions for similar cross sectional views. The data acquired by this arrangement provides a simple way for measurement and comparison of the performance of a system i.e., image resolution in terms of position and size of hot and cold regions in the phantom, region(s) detectability. Also the accuracy of scatter, attenuation correction and image reconstruction techniques in nuclear medicine tomography may be tested.
format Conference or Workshop Item
author Shah, Sayed Inayatullah
author_facet Shah, Sayed Inayatullah
author_sort Shah, Sayed Inayatullah
title PET/SPECT Test Phantom
title_short PET/SPECT Test Phantom
title_full PET/SPECT Test Phantom
title_fullStr PET/SPECT Test Phantom
title_full_unstemmed PET/SPECT Test Phantom
title_sort pet/spect test phantom
publishDate 2010
url http://irep.iium.edu.my/6054/
http://irep.iium.edu.my/6054/1/IRIIE_2010_Poster.pdf
first_indexed 2023-09-18T20:14:53Z
last_indexed 2023-09-18T20:14:53Z
_version_ 1777407721192030208